A&D FX-2000i Manual, Explained: Setup, Units and Calibration

The A&D FX-2000i manual, made simple: set up the 2200 g × 0.01 g precision balance, warm it up for 30 minutes, and calibrate with an external 2000 g weight. The FX-2000i is in stock at MNM Scales for $618.00 (regular $1,505.00).

Short answer: the official A&D FX-2000i manual is the FZ-i / FX-i Series Instruction Manual, part number 1WMPD4001863C. It covers installation, the keys, units, external calibration, the function table, and the specification table for the FX-2000i in the 1200i–5000i group. Place the balance on a stable bench, connect the AC adapter, and warm it up for at least 30 minutes with nothing on the pan (official sections 3.1, 3.4 and 8.5). Level the bubble, then calibrate with an external weight: the factory setting is 2000 g, and 1000 g and 500 g are also usable (section 8.5). Here is an NTEP alternative: the A&D FX-2000iN, 2200 g × 0.01 g, NTEP Class II, Certificate of Conformance 08-045A6 (Active). For how external calibration compares with an internal mass, see internal vs external calibration. Related FX-i family guides: FX-120i vs FX-120iN vs FX-300i, how to calibrate an A&D FX-120i, FX-120i AutoTrickler setup, and FX-123 vs FX-120i vs EJ-54D2. MNM Scales is an authorized A&D dealer and stocks the FX-2000i at $618.00 (regular $1,505.00).

We sell, calibrate and repair A&D FX-i balances from our Cypress, TX shop, and we ship from our Houston-area warehouse at 21249 FM 529 Rd, Cypress, TX 77433.

A&D FX-2000i manual at a glance: specifications

Figures below are from the official FZ-i / FX-i Series Instruction Manual, part number 1WMPD4001863C, section 20 specifications, FX-2000i column, and the section 8 calibration-weight table, read on November 18, 2026.

ItemFX-2000i
Capacity × readability2200 g × 0.01 g
Maximum display2200.84 g
Repeatability0.01 g
Linearity±0.02 g
Stabilization time, typical, FASTApprox. 1 s
Sensitivity drift, 10 °C to 30 °C±2 ppm/°C
Weighing panØ 150 mm
Net weight, approx.3.0 kg
Dimensions193 × 262.5 × 85.5 mm
InterfaceRS-232C
Operating environment5 °C to 40 °C, 85% RH or less, non-condensing
CalibrationExternal weight (factory setting 2000 g; 1000 g and 500 g also usable)
Warm-up timeAt least 30 minutes, nothing on the pan (sections 3.4 and 8.5)

Live prices at MNM Scales

Prices and stock are from our live store on November 18, 2026.

ModelCapacity × dCalibrationPriceStock
A&D FX-2000i (#510)2200 g × 0.01 gExternal, 2000 g factory setting$618.00 (regular $1,505.00)In stock
A&D FX-2000iN (#500)2200 g × 0.01 g, NTEP Class II, e = 0.1 gNTEP alternative, CC 08-045A6$799.00 (regular $1,555.00)In stock
A&D FX-2000iWP (#511)2200 g × 0.01 g, IP65External$970.00 (regular $1,955.00)In stock
A&D FX-2000iWPN (#42036)2200 g × 0.01 g, IP65, NTEP Class IINTEP washdown, CC 08-045A6$882.00 (regular $2,015.00)In stock

Setup from the A&D FX-2000i manual

Install the FX-2000i where temperature and humidity stay moderate (official section 3.1). Set the pan on the balance, level it so the bubble sits in the center of the spirit level, and connect the AC adapter. With the adapter connected the balance can sit in standby; that is a normal state. For accurate weighing, warm it up for at least 30 minutes before use, with nothing on the pan (sections 3.4 and 8.5). The MODE key cycles the weighing units stored in the balance; SAMPLE turns the minimum weighing value on or off in weighing mode.

How to calibrate the FX-2000i

The FX-2000i is calibrated with an external weight. Internal-mass one-touch calibration is the FZ-i series procedure (section 8.2). For the FX-2000i, official section 8.5 is the external-weight sequence:

  • Warm up the balance for at least 30 minutes with nothing on the pan.
  • Press and hold the CAL key until Calout appears, then release the key. The balance displays Cal 0.
  • To keep the stored 2000 g factory weight, continue. To choose 1000 g or 500 g, or to adjust the stored value, press SAMPLE and use RE-ZERO, MODE and PRINT as described in section 8.5.
  • Confirm the pan is empty and press PRINT. The balance measures the zero point. Keep the bench free of vibration and drafts while it measures.
  • Place the displayed calibration weight on the pan and press PRINT. The balance measures the weight.
  • When end appears, remove the weight. If GLP output is enabled, the balance can print a calibration report.

Section 8 also recommends a new calibration when the balance is installed for the first time, after it is moved, and when the ambient environment has changed. For a shorter walk-through of the same FX-i external-cal keys on a smaller model, see how to calibrate an A&D FX-120i. For a comparison of internal-mass and external-weight balances, see internal vs external calibration.

FX-2000i vs FX-2000iN

The FX-2000i and the FX-2000iN are both 2200 g × 0.01 g FX-i precision balances with external calibration. The FX-2000i is in stock at $618.00 (regular $1,505.00). The FX-2000iN is the legal-for-trade sibling, NTEP Class II, Certificate of Conformance 08-045A6 (Active), in stock at $799.00 (regular $1,555.00). On that certificate the FX-2000iN is 2,200 g, nmax 22,000, e = 0.1 g, d = 0.01 g, with a 150 mm platform, and the counting feature is labeled for prescription use only. The washdown IP65 models are the FX-2000iWP at $970.00 and the NTEP FX-2000iWPN at $882.00, also listed on CC 08-045A6. For how smaller FX-i models compare with each other, see FX-120i vs FX-120iN vs FX-300i.

Notes from our Cypress, TX shop

MNM Scales is an authorized A&D dealer. We keep the FX-2000i in stock in Cypress and can add in-house or NIST-traceable calibration before the balance ships. Our technicians also handle mail-in work from anywhere in the US and on-site service in the Houston area. The FX-2000i is a 2200 g precision balance; if you are setting up a smaller FX-i for reloading or an AutoTrickler, use the FX-120i guides linked above rather than this page’s 2000 g calibration weight.

NTEP and the FX-2000iN

The FX-2000i is a non-NTEP model for in-house use. Here is an NTEP alternative: the A&D FX-2000iN, 2200 g × 0.01 g, NTEP Class II, Certificate of Conformance 08-045A6 (Active, A&D Engineering Inc., FX Series, issued August 23, 2024). The same certificate also lists FX-2000i-P, FX-2000iWPN, FX-2000iWP-P and FX-2000GD. For definitions, use what NTEP and legal for trade mean.

Calibration, repair and rentals

Request calibration through Get a Quote or call (832) 290-3120, or start from our Houston scale calibration page. The 2200 g, 0.01 g FX-2000i sits on the 2.2 kg capacity boundary, which stays in the up-to-2.2 kg tier: in-house calibration with a certificate is $95.00, and NIST-traceable calibration with a certificate is $395.00, both in stock as add-on services. Mail-in calibration is nationwide. Calibration usually takes 1–3 business days after we receive the balance. On-site service is a $295.00 service call plus $2 per mile round trip, with calibration priced per scale on top. We repair all brands and types of scales, including A&D FX-i balances. For scale rentals, call (832) 290-3120 or request a quote.

Frequently asked questions

Where can I get the A&D FX-2000i manual?

The official A&D FX-2000i manual is the FZ-i / FX-i Series Instruction Manual, part number 1WMPD4001863C. This page walks through setup, the keys, external calibration, and the official specification table for the A&D FX-2000i we stock at $618.00 (regular $1,505.00).

How do I set up the A&D FX-2000i?

Place it on a stable bench where temperature and humidity stay moderate (official section 3.1). Fit the 150 mm pan, level the bubble in the spirit level, and connect the AC adapter. Warm the balance up for at least 30 minutes with nothing on the pan (sections 3.4 and 8.5), then calibrate with the external 2000 g weight (1000 g and 500 g are also usable) before the first accurate weighing.

How do I calibrate the A&D FX-2000i?

Use official section 8.5. Warm up the balance for at least 30 minutes with nothing on the pan. Press and hold CAL until Calout appears, then release the key. Keep the stored 2000 g factory weight, or press SAMPLE to select 1000 g or 500 g. Confirm the pan is empty and press PRINT to measure zero. Place the displayed weight on the pan and press PRINT. When end appears, remove the weight. Keep the bench free of vibration and drafts while the balance is measuring.

Is the A&D FX-2000i NTEP?

The FX-2000i is a non-NTEP model for in-house use. Here is an NTEP alternative: the A&D FX-2000iN, 2200 g × 0.01 g, NTEP Class II, Certificate of Conformance 08-045A6 (Active), in stock at $799.00. For definitions, see what NTEP and legal for trade mean.

How does the FX-2000i compare with the FX-2000iN?

Both are 2200 g × 0.01 g FX-i precision balances with external calibration. The FX-2000i is in stock at $618.00 (regular $1,505.00). The FX-2000iN is the legal-for-trade sibling, NTEP Class II, Certificate of Conformance 08-045A6 (Active), e = 0.1 g, d = 0.01 g, in stock at $799.00 (regular $1,555.00). IP65 models are the FX-2000iWP and the NTEP FX-2000iWPN.

Can I add calibration to an FX-2000i before it ships?

Yes. The 2200 g FX-2000i stays in the up-to-2.2 kg capacity tier: in-house calibration with a certificate is $95.00, and NIST-traceable calibration with a certificate is $395.00, both in stock as add-on services. Mail-in calibration is nationwide, and turnaround is usually 1–3 business days after we receive the balance.

About MNM Scales

MNM Scales has sold, calibrated and repaired all brands and types of scales for over 30 years. We are an authorized A&D dealer. Our TDA-registered technicians calibrate in-house with certificates, offer mail-in and on-site calibration, place commercial scales in service in Texas, and rent all types of scales. We supply customers across food, cannabis and hemp, compounding, jewelry, aerospace, oil and gas, and pawn shops. Shop: 21249 FM 529 Rd, Cypress, TX 77433 (by appointment). Call (832) 290-3120 or contact us.

Last updated November 18, 2026. Sources: official A&D FZ-i / FX-i Series Instruction Manual, part number 1WMPD4001863C, sections 3.1, 3.4, 8, 8.2, 8.5 and 20 (FX-2000i column); official NTEP Certificate of Conformance 08-045A6 (Active, A&D Engineering Inc., FX Series, issued August 23, 2024, including FX-2000iN and FX-2000iWPN); official NCWM API records previously verified for 08-045A6; live MNM Scales listings #510, #500, #511 and #42036. Prices from our live store on November 18, 2026.

Leave a Comment