Scale readability vs accuracy are not the same thing. Readability (resolution, often written as d) is the smallest increment the display can show. Accuracy is how close a reading is to the true mass — bounded by the instrument’s tolerance, calibration and uncertainty. A scale can read to 0.001 g and still be wrong by several milligrams if it is uncalibrated, off-level or used outside its class limits.
Definitions that hold up in metrology
| Term | Meaning | Typical symbol |
|---|---|---|
| Readability / resolution | Smallest difference between indicated values | d (scale division) |
| Verification scale interval | Division used for tolerance tests on legal-for-trade instruments | e (often e = d) |
| Accuracy (colloquial) | Closeness to the true value; in legal metrology, staying inside maximum permissible error | MPE / tolerance |
| Repeatability | Agreement of repeated indications under the same conditions | Often stated as σ |
| Linearity | How span error changes across the capacity | ± value in the datasheet |
Worked example: fine digits, coarse truth
A jeweler’s balance shows 10.004 g for a 10 g weight (d = 0.001 g). That does not prove 4 mg accuracy. If the balance’s Class II tolerance at that load is several milligrams, and the last calibration was months ago in a different room, the hundredths and thousandths can be noise relative to the real error. Conversely, a Class III shipping scale with d = 0.1 lb can be highly accurate for parcel work if it is NTEP, sealed and within maintenance tolerance — even though it will never show milligrams.
Multi-interval example: EW-1500i
The A&D EW-1500i is multi-interval: 0.1 g up to 300 g, 0.2 g up to 600 g, 0.5 g up to 1500 g. Readability gets coarser as load rises. That is by design under Handbook 44 multi-interval rules, not a loss of “accuracy class” by itself — the Class III certificate (CoC 04-019A3) sets the legal tolerances. Manual: EW-1500i manual explained.
How to choose readability without overbuying
- Process need: pick d about 1/5 to 1/10 of the smallest error you care about (rule of thumb, not a law).
- Weight class: calibration weights must support that d — see calibration weights.
- Environment: finer d needs better leveling, draft control and warm-up.
- Trade use: buy the NTEP class your commodity requires, not just more digits — NTEP guide.
Prove accuracy: calibrate and document
Accuracy is demonstrated, not assumed from the brochure. Use known test weights, record as-found/as-left, and recalibrate on a schedule. MNM Scales calibrates complete scales in Houston from $95 (≤2.2 kg in-house tier) with optional NIST-traceable certificates.
Related explainers
- How temperature, level and drafts affect scale accuracy
- Why do two scales show different readings?
- Scale readability vs accuracy
Frequently asked questions
What is the difference between accuracy and resolution on a scale?
Resolution (readability, d) is the smallest display step. Accuracy is closeness to the true value within the allowed tolerance after proper calibration and use.
Is a scale with more decimal places more accurate?
Not necessarily. Extra digits can display noise. Accuracy comes from design class, calibration, environment and tolerance — not from digit count alone.
What do d and e mean on a scale?
d is the scale division (display). e is the verification scale interval used for legal tolerances; on many instruments e equals d.