A&D EK-610i Manual, Explained: Setup, Units, Counting and 0.01 g Calibration
Short answer: the A&D EK-610i runs on the shared EK-i/EW-i instruction manual. For the EK-610i it comes down to six jobs: set it on a steady, draft-free bench and level it; warm it up for 30 minutes; store the units you use; count with a sample of at least 0.01 g per piece; keep the reference at 1 g or more in percent mode; and calibrate with a weight of at least 400 g (600 g is the factory value). Download the manual from A&D’s EK-610i page; this guide puts the EK-610i numbers in one place.
View the EK-610i: price and stock Call (832) 290-3120 Get a quote
Related A&D guides
- A&D EK-610i manual, explained (you are here)
- EK-1200i vs EK-610i: which one to buy
- How to calibrate an A&D EK-i balance (EK-1200i guide)
- A&D EK-1200i manual, explained
- The A&D EK-i series compared
- A&D EK-6100i manual, explained
- A&D EK-12Ki manual, explained
- A&D FG-60KBM manual, explained (our bench scale guide)
The A&D EK-610i is a 600 g x 0.01 g compact balance with a 110 mm round stainless pan, a backlit display and external calibration. A&D writes one instruction manual for the whole EK-i and EW-i family, so EK-610i owners have to pick their numbers out of shared tables. Below are the EK-610i figures from that manual, with the display messages exactly as the balance shows them.
EK-610i manual at a glance: specifications
| Item | EK-610i |
|---|---|
| Capacity x readability | 600 g x 0.01 g |
| Repeatability (standard deviation) | 0.01 g |
| Linearity | ±0.02 g |
| Sensitivity drift | ±20 ppm per °C (10 to 30 °C) |
| Stabilization time | About 2 seconds |
| Counting | Samples of 5, 10, 25, 50 or 100 pieces; up to 60,000 pieces; minimum unit weight 0.01 g |
| Percent mode | Minimum reference (100 percent) weight 1 g; readings to 0.1 percent |
| Display | Backlit 7-segment LCD, 16 mm characters, 10 updates per second |
| Pan and weight | 110 mm (4.3 in) round pan; balance about 1.1 kg |
| Power | AC adapter, or the EKW-09i rechargeable battery pack (not included) |
| Operating range | -10 to 40 °C (14 to 104 °F), under 85 percent relative humidity, no condensation |
| Calibration weight (factory setting) | 600 g, external; the value can be changed |
Setting up the EK-610i
- Check the box: balance, pan, pan support, AC adapter and instruction manual. The battery pack, printer, RS-232C cable and calibration weight are not included unless ordered. Keep the box and foam for mail-in service.
- Pick the spot: a firm, level bench away from drafts, vents, direct sun, vibration and magnetic fields. At 0.01 g even a ceiling fan shows up in the last digit.
- Let it acclimate: a balance that rode in a hot or cold truck should reach room temperature before you switch it on.
- Fit the pan support and round pan, then turn the leveling feet until the bubble in the spirit level is centered.
- Power it: plug in the AC adapter and press ON/OFF. The display shows all segments, then
0.00 g. - Warm up for 30 minutes, then calibrate; the steps are in the EK-610i calibration section below.
Keys and the function menu
| Key | When weighing | In the function menu |
|---|---|---|
| ON/OFF | Power on or off | – |
| MODE | Steps through the stored units, pcs and percent | Returns to weighing |
| SAMPLE | Starts sample storing in pcs or percent; hold to open func | Moves to the next item |
| Sends the reading to a printer or PC | Stores the setting | |
| RE-ZERO | Zeroes the display or tares a container | Changes the value |
Hold SAMPLE until func appears to reach settings such as Cond (response speed), st-b (stability band), trc (zero tracking), poff (auto power-off), ltUp (backlight), prt and bps (data output). The full menu table, item by item, is in our EK-1200i manual guide; the EK-610i menu is the same.
Weighing units on the EK-610i
| Unit | Display | EK-610i capacity x readability |
|---|---|---|
| Gram | g | 600 x 0.01 g |
| Ounce (avoir) | oz | 21.164 x 0.001 oz |
| Pound | lb | 1.3228 x 0.0001 lb |
| Troy ounce | ozt | 19.290 x 0.001 ozt |
| Metric carat | ct | 3000.0 x 0.1 ct |
| Momme | mom | 160.000 x 0.005 mom |
| Pennyweight | dwt | 385.81 x 0.01 dwt |
| Grain | GN | 9259.4 x 0.2 GN |
Store only the units you use: hold SAMPLE until func shows, press SAMPLE until Unit appears and press PRINT. Press SAMPLE to step to a unit and RE-ZERO to store it; repeat for each unit you use, then press PRINT (end, then id) and MODE to return to weighing. The balance starts with the first unit you stored. The units available on your balance are set at the factory. For conversions on paperwork, use our grain to gram converter.
Counting parts on the EK-610i
- Press MODE until
pcsshows, then press SAMPLE. - Press SAMPLE again to choose 5, 10, 25, 50 or 100 pieces.
- Set a container on the pan if you use one and press RE-ZERO so the display shows zero beside the sample size.
- Count that many pieces onto the pan and press PRINT to store the unit weight; remove them.
- Load the parts to count and read the piece count.
- On the EK-610i the smallest unit weight is 0.01 g and the maximum count is 60,000 pieces, so it suits small screws, seeds and labels.
- If the sample is light, the balance asks for more pieces with a prompt such as
25 -; add that many and press PRINT.lomeans the unit weight is too light to store. - ACAI (
aCai) refines the unit weight when you add a few more pieces and wait for the ACAI mark to go out; leave the pan alone while it blinks. - Counting is for inventory and production. For heavy volume counting, see our counting scale buying guide.
Percent mode and the HI/OK/LO comparator
Percent mode reads a load against a reference you store as 100 percent. Press MODE to the percent display, press SAMPLE, press RE-ZERO, place the reference and press PRINT. On the EK-610i the reference must weigh at least 1 g, and readings show to 0.1 percent. The comparator (Cp, CpHi, Cplo and bep in the function menu) shows HI, OK or LO against your limits for fill checks and kit checks.
RS-232C, printer and PC
- The port is a 9-pin D-Sub; use a straight-through cable to a PC (cable not included).
- Match
bps(baud rate) andbtpr(data bits and parity) to the other device, and choose the output withprt: on command, PRINT key, auto print on a stable load, or stream. - An A&D AD-8121 printer (not included) needs
pUseset to 1 for its 1.6-second pause. - For a USB-only computer, A&D lists the AX-USB-1920-9P USB to 9-pin RS-232 converter (not included). See connecting a scale to a computer by USB or RS-232.
Power: AC adapter and battery pack
- The EK-610i ships with its AC adapter; use the adapter supplied for the EK-i series.
- For cordless work add the EKW-09i rechargeable battery pack (not included). It fits the option slot on the back, the same slot the comparator relay option uses, so you fit one or the other.
- Charge the pack with the AC adapter connected and the balance off; a full charge takes about 15 hours, and a new pack may need two or three charges to reach full capacity.
- Battery run time depends on the backlight and use, so set
ltUpto a short timeout and turn onpoff. Recharge or plug in whenlb0appears. - Keep the adapter cord out of foot traffic; cut or pinched cords are a frequent repair. See replacing a scale cable.
EK-610i error messages
| Display | Meaning | What to do |
|---|---|---|
E | Overload (over 600 g) | Remove the load |
-E | Strong upward force on the sensor | Check for anything caught around the pan |
error5 | Unexpected load | Clear the pan and check that the pan and support are seated |
lo | Unit or 100 percent weight too light | Use more pieces (0.01 g minimum unit) or a reference of 1 g or more |
Cal E / -Cal E | Calibration weight too heavy / too light | Check the weight and the value you entered; press MODE |
error1 | Reading never settles | Stop drafts and vibration; press MODE |
lb0 | Battery pack discharged | Recharge or use the AC adapter |
Hb1 | AC adapter voltage too high | Use the correct EK-i adapter |
If error2, error3, error4 or error6 returns after a power cycle with an empty pan, send it to our repair bench. Other brands’ codes are in our scale error codes guide, and drifting readings in scale will not zero or keeps drifting.
EK-610i calibration: 600 g factory value, 400 g minimum
The EK-610i calibrates with an external weight. A&D calls for calibration when the balance is first used, after a move, when the room changes and on a regular schedule. The CAL switch is the push-button on the back behind a small access plate, the same as on the EK-1200i in our EK-i calibration guide.
- Warm up for at least 30 minutes with nothing on the pan.
- Hold the CAL switch until
Calappears, then release it; the display showsCal 0. - To use a weight other than 600 g, press SAMPLE: SAMPLE picks the digit, RE-ZERO sets it, PRINT stores the new value. The value stays in memory after power-off.
- With the pan empty, press PRINT to weigh zero; do not touch the pan. The display then shows the calibration value; to stop after the zero calibration only, press MODE here.
- Place the weight that matches the displayed value and press PRINT. When
endappears, remove the weight and press CAL or MODE to return to weighing.
A&D asks for a weight of more than 2/3 of capacity, so use at least 400 g on the EK-610i. A 600 g weight matches the factory value; our 500 g Class I calibration weight works after you set the value to 500.00 g. Our calibration weight classes guide explains which class suits a 0.01 g balance, and we stock calibration weights. Afterward, check a light load and one near capacity; our how often should a scale be calibrated guide helps set a schedule.
EK-610i vs EK-300i and EK-410i
| Model (our listing) | Capacity x readability | Max. count | Linearity | Factory calibration weight |
|---|---|---|---|---|
| EK-300i | 300 g x 0.01 g | 30,000 pcs | ±0.02 g | 300 g |
| EK-410i | 400 g x 0.01 g | 40,000 pcs | ±0.02 g | 400 g |
| EK-610i | 600 g x 0.01 g | 60,000 pcs | ±0.02 g | 600 g |
All three read 0.01 g, so the choice comes down to the heaviest container plus sample you weigh. Choosing between 0.01 g and an NTEP model for trade? Our EK-1200i vs EK-610i comparison covers that decision, and the EK-i series guide lists every model.
Selling by weight: choose the NTEP EK-1200i
The EK-610i is built for lab, production and inventory weighing at 0.01 g. A&D’s EK-i series holds NTEP Certificate of Conformance 04-019A4, listed as Active on NCWM’s certificate search on November 1, 2026; the EK-610i is not NTEP, as the certificate lists the EK-600i, EK-1200i, EK-6000i and EK-12Ki. For selling by weight, choose the NTEP Class II A&D EK-1200i (1200 g x 0.1 g). Our what is NTEP guide explains the approval, and our free legal-for-trade scale checker helps decide whether your use is trade.
Texas note: TDA does not require an NTEP certificate for every device, but the scale must meet NIST Handbook 44 and suit its use; some scales do need one, so we recommend NTEP models and can check yours. See TDA’s scales used in trade guidance (PDF) and our TDA scale certification guide, or call us or TDA to confirm your use.
Who uses the EK-610i
- Labs and schools: sample prep, formulation and teaching labs that need 0.01 g on a small footprint.
- Botanical and cannabis processors: in-house batch and moisture checks.
- Manufacturing and kitting: counting small parts down to 0.01 g each and HI/OK/LO fill checks.
- Hobby and craft benches: small batches and ingredients up to 600 g.
Setup tips from our technicians
- Our TDA-registered technicians recalibrate after every move, even across the room; at 0.01 g a new bench changes readings.
- Raise
Condone step under an air vent instead of chasing a wandering last digit. - Use a lightweight container and tare it; heavy glassware eats into the 600 g capacity.
- Handle calibration weights with gloves or tweezers so skin oil does not add mass.
- Wipe the pan with a damp cloth and mild detergent; skip solvents on the housing.
Common mistakes we see
- Calibrating with a 200 g weight; the EK-610i needs at least 400 g.
- Changing the calibration value and forgetting it stays in memory, so the next person uses the wrong weight.
- Counting tiny parts with a 5-piece sample and ignoring the
25 -prompt. - Hiding units in
Unitand then thinking the balance lost grams. - Weighing before the 30-minute warm-up is over.
Browse all A&D precision balances.
Ready to order? See current price and stock on the A&D EK-610i listing. Want it calibrated before it ships? Add in-house calibration with a certificate ($95) or NIST calibration with a NIST-traceable certificate ($395). Questions? Call (832) 290-3120 or send a quote request.
Calibration, repair and rentals from MNM Scales
- Book calibration: the EK-610i is in our up-to-2.2 kg tier: in-house calibration with a certificate is $95 and NIST calibration with a NIST-traceable certificate is $395. See our balance calibration service in Houston, all tiers on the scale calibration service page, and NIST vs in-house certificates.
- Mail-in calibration: pack the balance in its original box and foam and ship it with the mail-in calibration service.
- Cypress pickup by appointment: drop off or pick up at our Cypress, TX shop (21249 FM 529 Rd, Cypress, TX 77433).
- On-site service: on-site calibration is $295 plus $2 per mile, round trip, once per visit, plus the calibration fee for each scale. Our TDA-registered technicians bring certified test weights to labs, cannabis and botanical processors, schools and production lines across the Houston area.
- Repair: display, keypad, adapter and load cell problems go to our scale repair bench.
- Rentals: we rent all types of scales; ask about a scale rental while yours is in for service.
Frequently asked questions
Where can I download the A&D EK-610i manual?
A&D publishes the EK-i/EW-i instruction manual, which covers the EK-610i, on its EK-610i product page. This guide collects the EK-610i numbers from it.
How do I calibrate the EK-610i?
Warm it up for 30 minutes, hold the CAL switch until Cal appears and release it. At Cal 0, press PRINT with the pan empty, place the 600 g weight (or the value you set) and press PRINT. When end appears, remove the weight and press CAL or MODE.
What weight do I need to calibrate the EK-610i?
The factory value is 600 g. A&D asks for more than 2/3 of capacity, so any weight of at least 400 g works once you enter its value at Cal 0 with the SAMPLE key.
How do I change units on the EK-610i?
Press MODE to step through the stored units. To choose which units are stored, hold SAMPLE until func appears, go to Unit, press PRINT, turn units on or off with SAMPLE and RE-ZERO and store with PRINT.
What is the smallest part the EK-610i can count?
The minimum unit weight is 0.01 g at the factory setting, and the balance counts up to 60,000 pieces from samples of 5, 10, 25, 50 or 100.
Does the EK-610i come with a battery?
It ships with an AC adapter. The EKW-09i rechargeable battery pack is not included; it fits the option slot on the back and takes about 15 hours to charge fully.
Is the EK-610i NTEP approved?
The EK-610i is not on NTEP CC 04-019A4. For selling by weight, choose the NTEP Class II A&D EK-1200i.
What does Cal E mean on the EK-610i?
Cal E means the calibration weight was too heavy for the value set, and -Cal E means it was too light. Check the weight and the value, then press MODE and start again.