We are a US-based dealer shipping directly from Houston, TX!
Contact us for expedited, Alaska, Hawaii, and International shipping rates

We will BEAT competitors' pricing when you contact directly!

mnmscales@yahoo.com
(832) 290-3120

Contact Us for More Details on price beating competitors' when you text/email/phone call, and for payments, specific/niche models, bulk orders, use cases/hobbies, plus any other questions!
Direct Bank Transfers are accepted.

MNM Scales LLC
(832) 290-3120
mary@mnmscales.com
sales@mnmscales.com
21249 FM 529 Rd,
Cypress, TX 77433

We accept PayPal Buy Now, Pay Later feature. All credit/debit cards are accepted.

A&D FG-150KAL Calibration: CAL Switch Steps, 300 lb Weights and Gravity Setting

Short answer: to calibrate an A&D FG-150KAL, warm it up for 30 minutes with the pan empty, open the CAL switch cover on the rear of the display pod and hold the CAL switch until Cal 0 shows. With the pan empty and STABLE on, press MODE to set zero. Place 300 lb (or 150 kg, or the value you enter) and press MODE again; end appears. If you would rather have it done with certified weights and a certificate, book scale calibration with MNM Scales.

View the FG-150KAL: price and stock Call (832) 290-3120 Get a quote

The A&D FG-150KAL is a 300 lb bench scale on a 15.4 x 20.9 inch platform, and like every FG-K model it calibrates with an external weight through a CAL switch inside the display pod. This guide follows A&D’s FG-K instruction manual step by step, adds tips from our TDA-registered technicians, and shows when to hand the job to us. For the keys and settings, see the FG-150KAL manual guide.

When the FG-150KAL needs calibration

  • When the scale is first set up.
  • After it has been moved, even across the building.
  • When the room or environment changes, such as a move from an air-conditioned office to a dock.
  • On a regular schedule; our how often should a scale be calibrated guide suggests intervals by use.
  • When a known weight reads outside your tolerance.

What you need for FG-150KAL calibration

ItemDetails
Test weight300 lb or 150 kg is the factory value (full capacity). You can enter a smaller value; our technicians use certified weights at or near capacity so the whole range is checked.
Warm-upAt least 30 minutes powered on with nothing on the pan
Auto power-offSet F1 to 0, or leave something on the pan during warm-up, so the scale stays on
Unitlb or kg; calibration mode will not start in oz or pcs
ToolA screwdriver for the lock screws; press the CAL switch with a fingertip, since A&D warns a ballpoint pen can short the board
LocationCalibrate where the scale will be used, level and on a solid surface
Source: A&D FG-K instruction manual WM+PD4000966, section 8.

Weights in stacks are fine as long as the total matches the value on the display and every piece is certified. Our calibration weight classes guide explains which class suits a 300 lb bench scale, and we stock calibration weights.

FG-150KAL calibration step by step

  1. Open the CAL switch. Loosen the lock screws on the rear of the display pod and remove the CAL switch cover. The calibration switch sits on the board inside.
  2. Warm up the scale for at least 30 minutes with nothing on the pan, in lb or kg.
  3. Enter calibration. Press and hold the CAL switch until Cal 0 appears, then release it.
  4. Zero calibration. Make sure the pan is empty, wait for the STABLE indicator and press MODE. The scale stores zero and shows 5pn 1 with the span value, 300.00 in lb or 150.00 in kg.
  5. Change the value if needed. To use a different weight, press ZERO to pick the blinking digit and PRINT to raise it until the display matches your weight.
  6. Span calibration. Place the weight on the center of the pan, wait for STABLE and press MODE. end appears.
  7. Finish. Remove the weight and turn the scale off. Refit the CAL switch cover and tighten the lock screws.

To stop after the zero calibration only, turn the power off at 5pn 1. A value you change in step 5 is cleared at power-off, so the next calibration starts again from 300 lb or 150 kg.

Checking the result

Turn the scale on, let it settle and place the test weight again; it should read the weight value. Check a load near half capacity and one near the low end of your range, and try the weight at each corner of the platform for an even reading. Our scale tolerance calculator shows acceptable error for your division size. If readings drift or will not return to zero, see scale will not zero or keeps drifting.

Gravity acceleration setting

Gravity varies by location, and the FG-K series stores a gravity value to match. A&D’s manual lets you change it when a weight is not on hand after a move: hold the CAL switch until Cal 0, press PRINT to show the stored value (for example 9.798), change it with ZERO and PRINT to the value for your area from the gravity map at the back of the manual, and press MODE to return to Cal 0; turn the power off to finish, or continue to a weight calibration.

A calibration with a weight at the place of use makes the gravity setting unnecessary, and our technicians calibrate the FG-150KAL with certified weights, on site or in our shop.

FG-150KAL calibration in Houston: our service options

If you do not keep 300 lb of certified weights on hand, or an auditor or customer asks for a certificate, MNM Scales calibrates the FG-150KAL in the 35 to 150 kg tier. Book through our scale calibration service in Houston:

OptionWhat you getPrice
In-house calibrationCalibrated with NIST-tested weights, certificate includedSee the current rate on the listing
NIST calibrationCalibrated with a NIST-traceable certificateSee the current rate on the listing
On-site calibrationOur technicians come to you with test weights$295 plus $2 per mile, round trip, once per visit, plus the per-scale rate
Service prices from our service listings on October 31, 2026.

Mail-in works too: pack the scale in its original packing and use the mail-in calibration service, or drop it at our Cypress, TX shop by appointment. Our scale calibration cost guide and NIST vs in-house certificate guide help you choose.

Common calibration mistakes

  • Skipping the 30-minute warm-up that A&D calls for before calibration.
  • Starting in oz or pcs; switch to lb or kg first.
  • Pressing MODE before STABLE lights, at zero or at span.
  • Placing a weight that does not match the displayed value; the scale trusts the number on the display.
  • Stacking weights off center or touching the column; keep the load on the middle of the platform.
  • Calibrating in one room and using the scale in another.

Sealed scales used in trade

On the NTEP FG-150KALN, the CAL switch is sealed for commercial use: NTEP CC 05-101 describes a wire security seal through two thumb screws on the rear of the display. Keep that seal in place and call us; our legal-for-trade calibration service recalibrates commercial scales and places them in service in Texas. See our TDA scale certification guide and what is NTEP guide.

Texas note: TDA does not require an NTEP certificate for every device, but the scale must meet NIST Handbook 44 and suit its use; some scales do need one, so we recommend NTEP models and can check yours. See TDA’s scales used in trade guidance (PDF) and our TDA scale certification guide, or call us or TDA to confirm your use.

Ready to order? See current price and stock on the A&D FG-150KAL listing. Want it calibrated before it ships? Add in-house calibration with a certificate or NIST calibration with a NIST-traceable certificate in the 35 to 150 kg tier; see the current rate on each listing. Questions? Call (832) 290-3120 or send a quote request.

View price and stock Call (832) 290-3120 Request a quote

Frequently asked questions

How do I calibrate an A&D FG-150KAL?

Warm it up for 30 minutes, open the CAL switch cover on the rear of the display pod and hold the CAL switch until Cal 0 shows. With the pan empty press MODE, place the 300 lb (150 kg) weight or the value you set, and press MODE again. When end shows, remove the weight and turn the scale off.

Where is the CAL switch on the FG-150KAL?

On the board inside the display pod. Loosen the lock screws on the rear of the pod and remove the CAL switch cover to reach it. Press it with a fingertip, not a pen.

What weight do I need to calibrate the FG-150KAL?

The factory value is 300 lb or 150 kg, full capacity. You can enter a different value at the 5pn 1 step with ZERO and PRINT, and place a weight that matches it.

Why will my FG-150KAL not enter calibration mode?

The unit must be lb or kg. Switch away from oz or pcs with MODE, then hold the CAL switch until Cal 0 appears.

Do I need to set the gravity value?

Only if you cannot calibrate with a weight after a move. A calibration with a weight where the scale is used makes the gravity setting unnecessary.

How often should the FG-150KAL be calibrated?

On first setup, after every move, when the environment changes and on a regular schedule set by how heavily it is used and what your customers or auditors require.

Can MNM Scales calibrate my FG-150KAL?

Yes. It is in the 35 to 150 kg tier: in-house calibration with a certificate or NIST calibration with a NIST-traceable certificate (current rates on our calibration service page), by mail-in, drop-off in Cypress or on site. Call (832) 290-3120.

About the author: MNM Scales

MNM Scales has sold, calibrated and repaired scales and balances for over 30 years. Our TDA-registered technicians calibrate in-house with certificates, offer mail-in and on-site calibration, place commercial scales in service in Texas, and rent all types of scales. We supply customers across aerospace, oil and gas, food, plastics, chemicals, pharmacies to big pharma, and pawn shops. Shop: 21249 FM 529 Rd, Cypress, TX 77433 (by appointment). Call (832) 290-3120 or contact us.

Last updated October 31, 2026. Sources: A&D FG-K series instruction manual WM+PD4000966 (section 8), NTEP CC 05-101, our FG-150KAL listing, our calibration service page, and TDA scales used in trade guidance. Service prices from our service listings on October 31, 2026.

Leave a Comment