A&D HR-250AZ Manual: Internal Mass, One-Touch Calibration and Setup
This HR-250AZ manual page covers setup, one-touch internal calibration, external calibration, counting and error codes for the A&D HR-250AZ Galaxy analytical balance, a 252 g × 0.1 mg balance with a built-in 200 g mass that we have in stock at $1,349.00.
Short answer: the official HR-250AZ manual is A&D’s HR-AZ / HR-A Series instruction manual (PDF), P/N 1WMPD4002375, which we host. To set the balance up, fit the pan support, pan, breeze break ring and breeze break, level it with the feet until the bubble is centered, connect the AC adapter and warm it up for at least one hour with the pan empty. To calibrate, fit the large breeze break and press CAL: the balance runs one-touch calibration with its internal mass and shows end. The HR-250AZ is a Non-NTEP balance. Here is an NTEP alternative: Sartorius BCE224-1NUS, 220 g × 0.1 mg (e = 1 mg), NTEP Class I (CC 22-017A1, Active), $1,998.00 (regular $3,650.00), in stock. MNM Scales calibrates 0.1 mg readability balances in-house with a certificate for $95.00, NIST-traceable for $395.00, or on site for $295 plus $2 per mile round trip.
In stock: $1,349.00 (regular $3,195.00) for the A&D HR-250AZ. Contact us or call (832) 290-3120 with a question about setup, calibration or an order.
Analytical balance guides from MNM Scales
- A&D HR-250AZ manual: internal mass, one-touch calibration and setup (you are here)
- A&D HR-250A manual: the HR-A family and external calibration
- A&D HR-150AZ manual
- Analytical balance price guide
- Lab scale guide
- Calibration weights by capacity and readability
- Scale calibration cost by capacity
- What NTEP and legal for trade mean
This HR-250AZ manual guide follows A&D’s HR-AZ / HR-A Series instruction manual (PDF) for the A&D HR-250AZ, the 252 g Galaxy with a built-in 200 g mass for one-touch calibration. The same manual also covers the external-calibration HR-250A, which has its own guide, and the 152 g HR-150AZ. Here is what is specific to the HR-250AZ.
HR-250AZ at a glance
Prices are from our live store on November 26, 2026.
| Model | Capacity × readability | NTEP | Price |
|---|---|---|---|
| A&D HR-250AZ | 252 g × 0.1 mg, internal mass, one-touch calibration | Non-NTEP. Here is an NTEP alternative: Sartorius BCE224-1NUS, 220 g × 0.1 mg (e = 1 mg), NTEP Class I (CC 22-017A1, Active), $1,998.00 (regular $3,650.00), in stock. | $1,349.00 (regular $3,195.00), in stock |
| Sartorius BCE224-1NUS | 220 g × 0.1 mg (e = 1 mg), external calibration | NTEP Class I, CC 22-017A1 | $1,998.00 (regular $3,650.00), in stock |
| Item | HR-250AZ |
|---|---|
| Weighing capacity | 252 g (maximum display 252.0084 g) |
| Minimum display | 0.1 mg (0.0001 g) |
| Repeatability (standard deviation) | 0.1 mg from 0 to 200 g, 0.2 mg from 200 to 250 g |
| Linearity | ±0.3 mg |
| Stabilization time (typical at FAST) | About 2 seconds (about 3 seconds at MID.) |
| Sensitivity drift | ±2 ppm/°C (10 to 30 °C) |
| Internal mass | Yes, 200 g |
| Clock and calendar | Yes |
| External calibration weights | 250 g (factory setting), 200 g, 100 g or 50 g |
| Counting mode | Minimum unit mass 0.1 mg; 10, 25, 50 or 100 samples |
| Percent mode | Minimum 100% reference mass 10.0 mg |
| Interface | RS-232C |
| Weighing pan | 90 mm diameter |
| Operating environment | 5 to 40 °C (41 to 104 °F), 85% RH or less, no condensation |
| Dimensions and weight | 198 (W) × 294 (D) × 315 (H) mm, about 3.9 kg |
| NTEP | Non-NTEP. Here is an NTEP alternative: Sartorius BCE224-1NUS, 220 g × 0.1 mg (e = 1 mg), NTEP Class I (CC 22-017A1, Active), $1,998.00 (regular $3,650.00), in stock. |
Source: the A&D HR-AZ / HR-A Series instruction manual (PDF), section 21 (specifications) and sections 8-1 to 8-6 (calibration).
Setting up the HR-250AZ
- Place it: put the balance on a solid weighing table away from sunlight, heaters, air conditioners, drafts, vibration and magnetic fields. A&D recommends about 20 °C and 50% relative humidity.
- Assemble the pan: fit the pan support, weighing pan, breeze break ring and breeze break as shown in section 2-1. Pull out the locking handles, install the large breeze break and push the handles back in.
- Level it: turn the leveling feet until the bubble in the spirit level is centered. Poor leveling can cause a calibration error with the internal mass.
- Power it: check that the AC adapter type matches your voltage and receptacle, then connect it. The plug is dust-protected, so turn it while pushing.
- Warm up: leave the pan empty for at least one hour before calibrating or weighing. In standby the balance is simply on and warming, which does no harm.
The weighing pan edge is sharp. The breeze break parts have an antistatic coating, so do not wash them with water. HR-AZ balances can pick up static in dry air, so keep the humidity above 45% RH, wipe charged plastic samples with a moistened cloth, or use the optional AD-1683 DC static eliminator.
Keys and weighing
| Key | Press | Press and hold |
|---|---|---|
| ON:OFF | Turns the display on or off | – |
| SAMPLE | Turns the minimum weighing value on or off; stores a sample in counting or percent mode | Enters the function table |
| MODE | Switches the weighing units stored in the function table | Enters response adjustment |
| CAL | Cancels an operation. On the HR-AZ, starts one-touch calibration with the internal mass | Enters the calibration mode |
| Sends the weight to a printer or PC over RS-232C, and confirms settings | No function at the factory setting | |
| RE-ZERO | Sets the display to zero (tare) | – |
To weigh, place a container and press RE-ZERO, add the sample, wait for the stabilization indicator and read the value. Press PRINT while the indicator is on to output the result. Set the response to FAST for target weighing or SLOW in drafty rooms: hold MODE until RESPONSE shows, then press MODE to choose FAST, MID. or SLOW.
Units on the HR-250AZ
Press MODE to step through the units stored in the function table. Capacity in each unit on the HR-250AZ:
| Unit | Capacity | Minimum display |
|---|---|---|
| Gram | 252 | 0.0001 |
| Ounce (avoirdupois) | 8.88 | 0.00001 |
| Troy ounce | 8.10 | 0.00001 |
| Metric carat | 1260 | 0.001 |
| Momme | 67.2 | 0.0001 |
| Pennyweight | 162.0 | 0.0001 |
| Grain (UK) | 3889 | 0.002 |
| Tola (India) | 21.6 | 0.00001 |
| Messghal | 53.7 | 0.0001 |
The balance also offers milligram, four tael units, a density mode, a programmable unit, and counting and percent modes. Arrange them in the function table under 1Unit.
One-touch calibration with the internal mass
The HR-250AZ calibrates itself with a built-in 200 g mass, so no external weight is needed for routine calibration. A&D says to calibrate before each use and whenever the balance is first installed, moved or the environment changes.
- Connect the AC adapter and warm up for at least one hour with the pan empty.
- Check that the bubble is centered in the spirit level.
- Fit the large breeze break and press CAL.
- The balance shows Calin and calibrates with the internal mass. Keep drafts and vibration away.
- It shows end and returns to weighing. With GLP output set to 1 or 2, it also sends or stores a calibration report.
If one-touch calibration is blocked, the function switch for internal-mass calibration is set to inhibit; permit it in the function switch menu (section 9). To check the balance without changing it, hold CAL until CCin shows to run a calibration test with the internal mass. A&D lists the normal test value for the HR-250AZ as 200.0000 g.
External calibration and internal mass correction
Use an external weight to check or correct the internal mass, or to calibrate when you want a traceable reference. The HR-250AZ accepts 250 g (factory setting), 200 g, 100 g or 50 g, and the value can be adjusted within ±0.0150 g. A calibration weight is not included. We stock calibration weights, and our calibration weights guide explains classes.
- Warm up for at least one hour with the pan empty.
- Hold CAL until Calout shows, then release.
- Cal 0 shows. Press SAMPLE if you want to change the weight, then use RE-ZERO and MODE to pick it and PRINT to store it.
- With the pan empty, press PRINT to measure zero.
- Place the displayed weight on the pan and press PRINT.
- When end shows, remove the weight. Place it again to confirm the display is within ±2 digits.
The internal mass can drift with corrosion or aging, so A&D recommends checking it periodically. The HR-AZ series can correct it in two ways, both of which first need the internal mass value correction switch set to permit (the factory setting inhibits it): method 1 enters a correction value measured with an external weight, with a correction reference of 200 g on the HR-250AZ, and method 2 calibrates with an external weight and lets the balance correct and store the internal mass itself.
Counting and percent modes
- Press MODE until the counting mode (PCS) shows, then press SAMPLE to choose 10, 25, 50 or 100 samples.
- Place a container if needed and press RE-ZERO. The sample count appears.
- Place that many pieces on the pan, wait for the stabilization indicator and press PRINT to store the unit mass.
- Place the parts to count. Add a few more pieces to use ACAI, which refines the unit mass automatically.
Use samples with a unit mass of at least ten times the minimum weighing value, so 1 mg or more on this balance. If lo shows, the sample is too light. In percent mode press MODE until % shows, press SAMPLE, tare, place the 100% reference and press PRINT. The stored unit mass and reference survive a power loss.
Error codes
| Display | Meaning and fix |
|---|---|
| EC, E11 | Stability error: the balance cannot settle. Reduce vibration, drafts, temperature change, static and magnetic fields, or raise the response to SLOW. |
| EC, E17 | Internal mass error (HR-AZ only): the internal mass mechanism is not working. Repeat the operation; call for service if it persists. |
| EC, E20 / E21 | Calibration weight too heavy / too light. Check that the pan is installed and the weight value, then press CAL. |
| Overload | A load above 252 g is on the pan. Remove it. |
| Weighing pan error | The weight is too light. Check the pan is installed and calibrate. |
| Sample or unit mass error | The counting or percent sample is too light. Use more or heavier samples. |
| Clock battery error | The clock backup battery is empty. Reset the date and time; the weighing is unaffected. |
| EC, E00 to E07 | RS-232C communication errors. Check the command format, baud rate and parity. |
Clean the balance with a lint-free cloth moistened with warm water and mild detergent, never organic solvents. For service, contact us.
Calibration service from MNM Scales
Our TDA-registered technicians calibrate 0.1 mg readability balances like the HR-250AZ in our shop, by mail-in from anywhere in the country, or at your site:
| Service | Price |
|---|---|
| In-house calibration with a certificate (NIST-tested weights) | $95.00 |
| NIST calibration with a NIST-traceable certificate | $395.00 |
| On-site service call | $295 plus $2 per mile round trip, plus the per-scale rate |
Add either calibration when you order the A&D HR-250AZ and we calibrate it with its certificate. Our scale calibration cost guide lists every tier, and (832) 290-3120 books service.
NTEP and legal for trade
NCWM’s certificate search lists no NTEP Certificate of Conformance for the HR-250AZ, so the HR-250AZ we sell at $1,349.00 is a Non-NTEP balance for labs, QC and research. Here is an NTEP alternative: Sartorius BCE224-1NUS, 220 g × 0.1 mg (e = 1 mg), NTEP Class I (CC 22-017A1, Active), $1,998.00 (regular $3,650.00), in stock. Its NCWM Certificate of Conformance 22-017A1 (PDF) lists the BCE224-1NUS as a Class I balance with e = 0.001 g and d = 0.0001 g, and we host it on our site. Read more in our guide to what NTEP and legal for trade mean.
Rules for commercial use
For commercial use in Texas, TDA inspectors use NIST Handbook 44 as the minimum standard. We recommend NTEP models, which meet those Handbook 44 requirements, and our TDA-registered technicians can check yours. See TDA’s scales used in trade guidance (PDF) and our TDA scale certification guide, or call us or TDA to confirm your use.
Frequently asked questions
Where can I find the HR-250AZ manual?
The official manual is A&D’s HR-AZ / HR-A Series instruction manual, P/N 1WMPD4002375, which covers the HR-250AZ. We host it as a PDF. This page walks through the setup, one-touch calibration and counting steps that apply to the HR-250AZ.
How do I calibrate the HR-250AZ?
Warm the balance up for at least one hour with the pan empty, check that it is level, fit the large breeze break and press CAL. It shows Calin, calibrates with its internal 200 g mass, then shows end and returns to weighing.
What weight do I need for HR-250AZ calibration?
Routine calibration uses the internal mass, so no weight is needed. For an external calibration or to correct the internal mass, use 250 g (factory setting), 200 g, 100 g or 50 g. We stock calibration weights.
What is the difference between the HR-250AZ and the HR-250A?
Both are 252 g × 0.1 mg Galaxy balances. The HR-250AZ has an internal mass with one-touch calibration and a clock and calendar. The HR-250A calibrates with an external weight only; see our HR-250A manual guide.
How do I count parts on the HR-250AZ?
Press MODE until the counting mode shows, press SAMPLE to pick 10, 25, 50 or 100 samples, tare a container, place that many pieces and press PRINT to store the unit mass. Then add the parts to count.
Is the HR-250AZ NTEP?
The HR-250AZ is a Non-NTEP balance for labs, QC and research. Here is an NTEP alternative: Sartorius BCE224-1NUS, 220 g × 0.1 mg (e = 1 mg), NTEP Class I (CC 22-017A1, Active), $1,998.00 (regular $3,650.00), in stock.
What does EC, E17 mean on the HR-250AZ?
EC, E17 is an internal mass error, which only the HR-AZ series can show. The mechanism that applies the internal mass is not working correctly. Repeat the operation, and call us if it keeps appearing.
How much does HR-250AZ calibration cost at MNM Scales?
For balances with 0.1 mg readability, in-house calibration with a certificate is $95.00 and NIST calibration with a NIST-traceable certificate is $395.00. On-site calibration is $295 plus $2 per mile round trip, plus the per-scale rate. Mail-in works nationwide.
About the author: MNM Scales
MNM Scales has sold, calibrated and repaired scales and balances for over 30 years. Our TDA-registered technicians calibrate in-house with certificates, offer mail-in and on-site calibration, place commercial scales in service in Texas, and rent all types of scales. We supply customers across aerospace, oil and gas, food, plastics, chemicals, pharmacies to big pharma and pawn shops. Shop: 21249 FM 529 Rd, Cypress, TX 77433 (by appointment). Call (832) 290-3120 or contact us.
Last updated November 26, 2026. Sources: A&D HR-AZ / HR-A Series instruction manual P/N 1WMPD4002375 (PDF) and NCWM Certificate of Conformance 22-017A1, both hosted on mnmscales.com; live store listings #1276, #1752, #42460 and #42461; and Texas Department of Agriculture legal-for-trade guidance. Prices from our live store on November 26, 2026.