OHAUS PX323 Manual: Setup, Calibration and Applications

This PX323 manual page covers setup, calibration, weighing applications and troubleshooting for the OHAUS Pioneer PX323, a 320 g × 0.001 g precision balance with internal calibration that we have in stock at $1,766.00.

Short answer: the official PX323 manual is the OHAUS Pioneer PX Series instruction manual (PDF, English section), which covers the PX323 and the rest of the PX line and which we host. To set the balance up, level it with the four feet until the bubble is centered, connect the AC adapter and let it acclimatize (1.5 hours for a balance at this precision). To calibrate, hold Tare / Menu-Cal, press Yes, and run Internal Cal, or run a span calibration with a 200 g or 300 g weight. The PX323 is a Non-NTEP balance. Here is an NTEP alternative: A&D GX-303AN, 320 g × 0.001 g (e = 0.01 g), internal calibration, NTEP Class II (CC 21-045A3, Active), $1,439.00 (regular $2,475.00), in stock. MNM Scales calibrates balances up to 2.2 kg in-house with a certificate for $95.00, NIST-traceable for $395.00, or on site for $295 plus $2 per mile round trip.

In stock: $1,766.00 (regular $2,084.00) for the OHAUS Pioneer PX323. Contact us or call (832) 290-3120 with a question about setup, calibration or an order.

Shop the PX323   Call (832) 290-3120   Get a Quote

This PX323 manual guide follows the OHAUS Pioneer PX Series instruction manual (PDF) for the OHAUS Pioneer PX323, the internal-calibration (InCal) model. The same manual covers the external-calibration (ExCal) PX323/E and the other PX balances; for the analytical models read our PX124 manual guide and PX224 manual guide.

PX323 at a glance

Prices are from our live store on November 26, 2026.

ModelCapacity × readabilityNTEPPrice
OHAUS Pioneer PX323320 g × 0.001 g, internal calibration (InCal)Non-NTEP. Here is an NTEP alternative: A&D GX-303AN, 320 g × 0.001 g (e = 0.01 g), internal calibration, NTEP Class II (CC 21-045A3, Active), $1,439.00 (regular $2,475.00), in stock.$1,766.00 (regular $2,084.00), in stock
OHAUS Pioneer PX323/E320 g × 0.001 g, external calibration (ExCal)Non-NTEP. Here is an NTEP alternative: A&D GX-303AN, 320 g × 0.001 g (e = 0.01 g), internal calibration, NTEP Class II (CC 21-045A3, Active), $1,439.00 (regular $2,475.00), in stock.$1,457.00 (regular $1,721.00), in stock
A&D GX-303AN320 g × 0.001 g (e = 0.01 g), internal calibrationNTEP Class II, CC 21-045A3$1,439.00 (regular $2,475.00), in stock
ItemPX323
Capacity320 g
Readability0.001 g
Repeatability (standard deviation)0.0008 g up to 5% of full load, 0.001 g above that
Linearity±0.002 g (typical ±0.0006 g)
Typical stabilization time2 seconds
Sensitivity temperature drift±3 ppm/K
Typical minimum weight (USP, 0.10%)1.6 g
Span calibration masses200 g or 300 g (not included)
Linearity calibration masses0, 150 g and 300 g
ApplicationsWeighing, parts counting, percent weighing, dynamic weighing, density determination, formulation
Display2-line backlit LCD
Weighing pan120 mm diameter, stainless steel
InterfacesRS232 and USB
PowerAC adapter, 100 to 240 V input, 12 VDC 0.5 A output
Operating conditions+10 to +30 °C for ordinary lab use (operable +5 to +40 °C), 80% relative humidity maximum
Dimensions and weight (draft shield model)209 × 321 × 309 mm (W × D × H), about 4.5 kg (10 lb)
NTEPNon-NTEP. Here is an NTEP alternative: A&D GX-303AN, 320 g × 0.001 g (e = 0.01 g), internal calibration, NTEP Class II (CC 21-045A3, Active), $1,439.00 (regular $2,475.00), in stock.

Source: the OHAUS Pioneer PX Series instruction manual (PDF), tables 9-1 and 9-2 (specifications), section 5.2 (calibration) and section 8.3 (troubleshooting).

Setting up the PX323

  1. Unpack it: the box holds the balance, the power adapter with its plug, the stainless steel pan and a warranty card.
  2. Place it: choose a location away from heat sources, rapid temperature changes, air currents and vibration, with enough space around it.
  3. Level it: turn the four leveling feet until the bubble in the small round window beside the display is centered in the circle. Level it again whenever you move it.
  4. Power it: plug the DC connector into the rear receptacle, then plug the AC adapter into an outlet.
  5. Acclimatize it: OHAUS suggests waiting before use. For a balance at this precision the acclimatization time is 1.5 hours.

The balance has two data ports on the rear, RS232 for a computer or printer with a straight-through serial cable and USB for a computer with a USB 2.0 Type A to Type B cable.

Keys and weighing

KeyShort pressPress and hold
On / ZeroTurns the balance on, or sets zeroOff
PrintSends the displayed value to the serial portUnit: changes the weighing unit
FunctionDepends on the applicationMode: changes the application mode
TareTares the balanceMenu-Cal: opens the main menu, with Calibration first
Yes / No / Back / ExitAccept, reject or raise a value, go back or lower a value, leave the menu

To weigh, press Zero, place the object on the pan and read the display once the stable asterisk shows. To tare, press Zero with an empty pan, place a container, press Tare and add the sample; the net weight shows. After you remove the container the display goes negative, and Tare clears it.

Calibrating the PX323

OHAUS offers three calibration methods on an InCal model: internal calibration, span calibration and linearity calibration. Do not disturb the balance during any of them. The PX323 is an InCal model, so it calibrates with its built-in mass, and you can also calibrate it manually with external masses. We stock calibration weights, and our calibration weights guide explains classes. Our internal vs external calibration guide explains the difference.

Internal calibration

  1. Make sure the balance is on, has warmed up to operating temperature, is level and has no load on the pan.
  2. Press and hold Tare / Menu-Cal, press Yes at the Calibration menu and choose Internal Cal.
  3. The screen shows the status. When it finishes, press any button to return to the application.

Span calibration

  1. With the pan empty, press and hold Tare / Menu-Cal and press Yes to enter the Calibration menu.
  2. Press No until Span Cal shows and press Yes.
  3. The display shows Place weights and the mass. Place the 300 g weight, which gives the best accuracy, or press Function / Mode to switch to the 200 g point and place that weight.
  4. Remove the weight. Calibration done shows when it is finished.

Linearity calibration

  1. With the pan empty, open the Calibration menu as above and press No until Linearity Cal shows, then press Yes.
  2. The balance captures zero, then prompts for the next weight. Place 150 g, remove it when prompted, then place 300 g.
  3. Linearity done shows when it is finished.

InCal Adjust can fine-tune the internal calibration by up to 100 divisions: place a test mass equal to the span value on the pan, note the difference in divisions, and enter it in the InCal Adjust menu. Perform an internal calibration first.

Counting, percent weighing and more

Counting parts

  1. Press Tare or Zero if necessary, then press and hold Function / Mode until Parts Counting appears and press Yes.
  2. At Clear APW?, press No to keep the last average piece weight, or Yes to clear it.
  3. With Sample size 10 flashing, press Yes to confirm, or press No or Back to raise or lower the sample size first (1 to 1000).
  4. Place that many pieces on the pan and press Function / Mode to set the average piece weight. The display shows the piece count.
  5. Add more pieces and the count updates. Press Function / Mode to view the piece weight or the total weight.

The minimum piece weight should be no less than 0.1 d, which is 0.0001 g on the PX323. APW Optimization recalculates the piece weight automatically as parts are added, when you add between one and three times the number already on the pan.

Percent weighing

Press and hold Function / Mode until Percent Weighing appears and press Yes. At Clear reference?, press Yes, place the reference sample on the pan and wait for the stable asterisk. The reference weight must be at least 100 d. Later samples then show as a percentage of the reference.

The PX also offers dynamic weighing, density determination and formulation; the manual walks through each one.

Troubleshooting

SymptomPossible causeRemedy
Cannot change menu settingsThe sub-menu is locked.Unlock it in the Lockout menu.
Parts count errorThe average piece weight is below 0.1 d.Re-evaluate the average piece weight or use a different sample.
Percent weighing errorThe reference weight is below 100 d.Put more samples on the pan.
Unstable or inaccurate readingsThe environment is not stable, or incorrect calibration masses were used.Move the balance to a suitable location and calibrate again with the masses in the specification table.
Pan error at power-onA wrong pan is fitted, the pan is not installed or the pan is not seated, or there is a load on the pan.Use the original OHAUS pan, install it properly and remove any load before power-on.
OverloadThe weight on the pan is too heavy.Reduce the sample until it is within the 320 g capacity.
System errorAn internal fault.Contact service if it persists.

Calibration service from MNM Scales

Our TDA-registered technicians calibrate balances up to 2.2 kg like the PX323 in our shop, by mail-in from anywhere in the country, or at your site:

ServicePrice
In-house calibration with a certificate (NIST-tested weights)$95.00
NIST calibration with a NIST-traceable certificate$395.00
On-site service call$295 plus $2 per mile round trip, plus the per-scale rate

Add either calibration when you order the OHAUS Pioneer PX323 and we calibrate it with its certificate. Our scale calibration cost guide lists every tier, and (832) 290-3120 books service.

Is the PX323 NTEP?

No. NTEP certificate 18-110A3 covers the N versions of the Pioneer, not this standard PX323. The PX323 we sell at $1,766.00 is a Non-NTEP balance for labs, QC and education. Here is an NTEP alternative: A&D GX-303AN, 320 g × 0.001 g (e = 0.01 g), internal calibration, NTEP Class II (CC 21-045A3, Active), $1,439.00 (regular $2,475.00), in stock. Its NCWM Certificate of Conformance 21-045A3 (PDF) lists the GX-303AN by model, and we host it on our site. Read more in our NTEP balance guide and guide to NTEP and legal for trade.

Rules for commercial use

For commercial use in Texas, TDA inspectors use NIST Handbook 44 as the minimum standard. We recommend NTEP models, which meet those Handbook 44 requirements, and our TDA-registered technicians can check yours. See TDA’s scales used in trade guidance (PDF) and our TDA scale certification guide, or call us or TDA to confirm your use.

Frequently asked questions

Where can I find the PX323 manual?

The official manual is the OHAUS Pioneer PX Series instruction manual, which covers the PX323 and the rest of the PX line. We host the English section as a PDF. This page walks through setup, calibration, applications and troubleshooting.

How do I set up the OHAUS PX323?

Level the balance with the four feet until the bubble is centered, connect the DC connector and the AC adapter, and let it acclimatize for 1.5 hours. Then calibrate it before use and after every move.

How do I calibrate the PX323?

With the pan empty, hold Tare / Menu-Cal, press Yes at the Calibration menu and choose Internal Cal. The balance applies its built-in mass and shows the status; press any button when it finishes. You can also run a span or linearity calibration with external weights.

What weights do I need for PX323 span and linearity calibration?

Span calibration uses 300 g, or 200 g at the half-capacity point. Linearity calibration uses 150 g and 300 g after the zero point. We stock calibration weights.

How do I count parts on the PX323?

Hold Function / Mode until Parts Counting shows and press Yes, keep or clear the average piece weight, confirm the sample size (10 by default), place that many pieces and press Function / Mode. Then add the parts to count.

Is the PX323 NTEP?

NTEP certificate 18-110A3 covers the N versions of the Pioneer, not this standard PX323. The PX323 is a Non-NTEP balance for labs, QC and education. Here is an NTEP alternative: A&D GX-303AN, 320 g × 0.001 g (e = 0.01 g), internal calibration, NTEP Class II (CC 21-045A3, Active), $1,439.00 (regular $2,475.00), in stock.

What is the difference between the PX323 and the PX323/E?

Both are 320 g × 0.001 g Pioneer PX balances covered by the same manual. The PX323 has internal calibration, while the PX323/E is calibrated with external weights only. The PX323/E is in stock at $1,457.00.

How much does PX323 calibration cost at MNM Scales?

For balances up to 2.2 kg, in-house calibration with a certificate is $95.00 and NIST calibration with a NIST-traceable certificate is $395.00. On-site calibration is $295 plus $2 per mile round trip, plus the per-scale rate. Mail-in works nationwide.

About the author: MNM Scales

MNM Scales has sold, calibrated and repaired scales and balances for over 30 years. Our TDA-registered technicians calibrate in-house with certificates, offer mail-in and on-site calibration, place commercial scales in service in Texas, and rent all types of scales. We supply customers across aerospace, oil and gas, food, plastics, chemicals, pharmacies to big pharma and pawn shops. Shop: 21249 FM 529 Rd, Cypress, TX 77433 (by appointment). Call (832) 290-3120 or contact us.

Last updated November 26, 2026. Sources: OHAUS Pioneer PX Series instruction manual (PDF) and NCWM Certificates of Conformance 18-110A3 and 21-045A3, all hosted on mnmscales.com; live store listings #2484, #1300, #2045, #42460 and #42461; and Texas Department of Agriculture legal-for-trade guidance. Prices from our live store on November 26, 2026.